Valco Instruments Co. Inc.
D3.P65
Rev. 9/07
Printed in USA
Pulsed Discharge
Detector Models
D-3-I-HP and
D-3-I-7890
Instruction Manual
® ®
Europe, Asia, and Africa contact:
tel: + 41 41 925-6200
fax: + 41 41 925-6201
VICI AG International
North America, South America, and Australia/Oceania contact:
tel: 800 367-8424
fax: 713 688-8106
Valco Instruments Co. Inc.
Vue de la page 0
1 2 3 4 5 6 ... 24 25

Résumé du contenu

Page 1 - Instruction Manual

Valco Instruments Co. Inc.D3.P65Rev. 9/07Printed in USAPulsed DischargeDetector ModelsD-3-I-HP andD-3-I-7890Instruction Manual® ®Europe, Asia, and Afr

Page 2

6InstallationGeneral Precautions •Do not use plastic/polymer or copper tubes for gas handling andinterconnectons. Use only stainless steel tubing wi

Page 3 - Table of Contents

7Agilent Technologies7890AGC SystemMODEL D3DETECTORFigure 2: A D3 detector in the “Back” position on a 7890ACAPTIVE SCREWS (4)COLUMN INLETELECROMETER

Page 4

8InstallationCAUTION: A D-3-I-HP will not work on a 7890 GC, and aD-3-I-7890 will not work on a 6890. If you are not surewhich model you have, call t

Page 5 - Introduction

9Installation6. Press the <Config> key, then press <Back det> or <Front Det>, corre-sponding to the location where the PDD will be i

Page 6 - Safety Notes and Information

10Installation11. Scroll to select the second line and press <Enter>. The display will read:CAUTION: Instrument power must be turned offand bac

Page 7 - Components of the D3 system

11InstallationGas ConnectionsRemember these three points discussed earlier: (1) all surfaces that contactthe gas stream must be glass or stainless ste

Page 8 - System Requirements

127. Open the cylinder valve to pressurize the regulator once again. Close thevalve and observe the needle of the high pressure gauge for 15 minutes.

Page 9 - Gas Specifications

13Initial Power-UpCAUTION: Always make sure that discharge gas isflowing before heating and powering up the detector.1. Before installing the column,

Page 10 - Installation

14Capillary Column ConnectionIf the capillary column adapter is installed in the column inlet:1. Make a mark on the column 10.3 cm from the end.2. Rem

Page 11

153. Set aside the 1/8" wrench and completely remove the adapter from thecolumn inlet.4. Screw the packed column adapter into the column inlet by

Page 12 - Agilent 7890 GC

This page intentionally left blank for printing purposes

Page 13

16Mode Selection and SetupHelium Ionization ModeSince the PDHID mode provides a better indication of the cleanliness andthe integrity (leak tightness)

Page 14

17Mode Selection and SetupAr-PDPIDChanging the discharge gas from helium to a mixture of 2% argon inhelium changes the photon energy level from the 17

Page 15 - Gas Connections

18TroubleshootingHigh Background CurrentLow SensitivityPurple/bluePinkDisconnect the columnand cap the column inletBaseline current is still high(>

Page 16

19High Noise LevelIf the noise lievel is high:1. See if it improves with the GC fan turned off.2. Check the column insertion position (10.3 cm for cap

Page 17 - Initial Power-Up

20WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyermay also have other rights that vary from state to state. For a peri

Page 18 - Packed Column Connection

21Detector Performance LogIn addition to the occasions indicated in the Comments area of the table below (see Initial Power-Up,page 13), the standing

Page 19 - Testing for Leaks

Table of ContentsIntroductionDescription and Operating Principles ... 1Safety Notes and Infor

Page 20 - Mode Selection and Setup

This page intentionally left blank for printing purposes

Page 21

1IntroductionDescription and Operating PrincipleThe D3 is a non-radioactive pulsed discharge ionization detector (PDID), withmodels available for “plu

Page 22 - Troubleshooting

2Safety Notes and InformationSymbolsATTENTIONRefer to the manual.Installation CategoryThis equipment has been designed for installation category (over

Page 23 - High Noise Level

3Components of the Detector SystemComponents of the detector system are listed in Table 1 below. Check thecontents of the packages to verify that eve

Page 24 - Warranty

4System RequirementsComponents Not Included with the Detector System • Helium (99.999% purity) and other support gases • Ultra high purity grade gas

Page 25 - Detector Performance Log

5IntroductionedoMrotceteDDIHDPDIPDP-rADIPDP-rKDIPDP-eXsagegrahcsiDmuileHeHnirA%2eHnirK%5.1eHnieX%8.0sagreirraCmuileH*****Gas Specifications** Any gas

Modèles reliés D-3-I-HP

Commentaires sur ces manuels

Pas de commentaire

PYLE Audio Altavoces del coche manuales de usuario

Empezando en M

Haga clic en el alfabeto a continuación para desplazarse a la lista completa de modelos que comienzan con esa letra

Modelos Tipo de documento
Mobile Audio System PLDTF5M
Manual de usuario   PYLE Audio Mobile Audio System PLDTF5M User Manual, 16 paginas
Mobile Audio System PLMR18
Manual de usuario   PYLE Audio Mobile Audio System PLMR18 User Manual, 14 paginas
Mobile Video System PLD41MUT
Manual de usuario   PYLE Audio Mobile Video System PLD41MUT User Manual, 26 paginas